Model 608
Model 608-II Control cabinet
Model 608-II 400l with round test chambers
Model 608-II 400l with rectangular test chambers
Cyclic corrosion tests, salt spray tests, condensation tests, in standard version for two test chambers with 400-/1000-/2000-litre effective volume
Corrosion Testing Equipment

608-2

Order number: 20550031
Non-Binding Inquiry
With the button below you can add products to your product inquiry list if you want us to do a quote for this product.
Note this product

Corrosion Testing Equipment

608-2

The Erichsen Model 608-2 (for two test chambers) is a compact corrosion tester for cyclic tests and alternating climate stress or alternating corrosion stress. The integrated touch control panel allows easy operation and monitoring of the device. Cyclic corrosion tests like alternating periods of salt fog, humidity and drying tests can be started in a user-friendly way and are then carried out fully automatically. An integrated storage tank stores the spray solution.

Salt spray chamber: Various test chamber sizes with test volume of 400 l, 1000 l or 2000 l can be selected regarding the requirements.
The test chambers (400 l and 1000 l) are available in circular and rectangular design. (For model 608-2: adapter for two salt spray chambers)

Purpose and application: Ferrous and non-ferrous metals are attacked continuously by humidity, acids, solutions, gases etc. It is therefore vitally important to choose the correct surface protection. There are many materials and qualities on the market and their properties must be properly assessed. Materials intended to prevent corrosion must be tested if failures are to be avoided. Furthermore the comparative quality control during production is of increasing importance. The best known processes employ spray vapour tests using various salt solutions as well as condensation water climates.

Test principle: Aggressive solutions are turned into a vapour mist in accordance with the tests that are listed below. These vapours surround the specimens in the test chamber either continuously or in a cyclic manner. The corrosion resistance of the individual specimens is established on the basis of the difference in time before the first corrosive effects become apparent.

Scope of delivery:

  • Control cabinet 608-2
  • Integrated solution reservoir
  • 3 specimen holders for each test chamber
  • 2 condensate receptacles for each test chamber
  • Manufacturer's Test Certificate

Norms

  1. DIN EN ISO 6270-2 CH
  2. DIN 50958
  3. DIN 55991
  4. ISO 11503
  5. ASTM D2247
  6. BS 3900 Part F2
  7. DIN EN ISO 6270-2 AHT
  8. DIN EN ISO 6270-2 AT
  9. ASTM B117
  10. D171058
  11. ISO 9227
  12. DIN EN ISO 9227 SS
  13. DIN EN ISO 9227 ESS
  14. DIN EN ISO 9227 CASS
  15. P-VW 1210

Technical specifications

Test 1

Condensation test

Test 2

Salt spray (fog) tests

Test 3

Cyclic corrosion test

Test 4

"Dry heat"

Effective Volume

400/1000/2000l

Power supply

400 V / 3AC, N, PE 50 Hz (other voltages on request)

Consumption (Single chamber)

max. 4.5 kVA

Consumption (Double Chamber)

max. 8.5 kVA (depending of the equipment)

Compressed air connection (Air pressure)

5 - 7 bar

Compressed air connection (Air consumption during ventilation)

15 l/min at 6 bar

(VE)Water connection Pressure

2 - 6 bar

Test temperature range

from ambient temperature up to +50 °C

Floor load of the test chamber

max. approx. 300 kg (special versions on request )

Capacity of the test chamber (400 l test chamber)

approx. 100 test panels

Capacity of the test chamber (1000 l test chamber)

approx. 180 test panels

Capacity of the test chamber (2000 l test chamber)

approx. 400 test panels (depending on the shape)

Equipment